AJNR Am J Neuroradiol 24:463–471, March 2003
Aneurysm Clips: Evaluation of Magnetic Field Interactions and Translational Attraction by Use of “Long-Bore” and “Short-Bore” 3.0-T MR Imaging Systems
Frank G. Shellock, Jean A. Tkach, Paul M. Ruggieri,
Thomas J. Masaryk, and Peter A. Rasmussen
BACKGROUND AND PURPOSE: The use of 3.0-T MR systems is increasing worldwide. We evaluated magnetic field interactions and translational attraction for 32 aneurysm clips in association with exposure to “long-bore” and “short-bore” 3.0-T MR imaging systems. METHODS: Thirty-two different aneurysm clips were evaluated in this investigation. Each aneurysm clip was qualitatively evaluated for magnetic field interactions and quantitatively assessed for translational attraction by using the deflection angle test. The deflection angle tests were performed at the points of the highest spatial gradients for long-bore and short-bore 3.0-T MR imaging systems. RESULTS: Seventeen of the 32 aneurysm clips showed positive magnetic field interactions. Deflection angles for the aneurysm clips were significantly (P < .001) higher on the short-bore (range, 0 –18 degrees) compared with those recorded on the long-bore (range, 0 –16 degrees) 3.0-T MR imaging system. Aneurysm clips made from commercially pure titanium and titanium alloy displayed no translational attraction (n ؍ 15), whereas those made from stainless steel alloy, Phynox, and Elgiloy displayed positive deflection angles (n ؍ 17). CONCLUSION: The 32 different aneurysm clips passed (angle <45 degrees) the deflection angle test by using the long- and short-bore 3.0-T MR imaging systems, indicating that they are safe for patients and other persons in these MR environments (ie, immediate area of MR imaging systems). However, only clips made from the titanium and titanium alloy are entirely safe for patients undergoing MR imaging procedures because of the total lack of magnetic field interactions. The remaining clips require characterization of magnetic field–induced torque. Because of possible differences in the points of the highest spatial gradients for different 3.0-T MR imaging systems, the results are specific to the imaging units and bore designs used in this investigation.
Neurosurgical management of an intracranial aneu-
tracranial aneurysm clip in a patient or other person
rysm or arteriovenous malformation by application of
in the MR environment may present a hazardous
a temporary or permanent aneurysm clip is a well-
situation (5–10). Although certain aneurysm clips are
established procedure (1–4). The presence of an in-
a contraindication to the MR environment, others
that are classified as “nonferromagnetic” or “weakly
ferromagnetic” are deemed safe for patients or other
Received April 25, 2002; accepted after revision July 2.
persons exposed to MR imaging systems operating at
Supported by The Cleveland Clinic Foundation, Cleveland, OH,
and the Institute for Magnetic Resonance Safety, Education, and
The use of 3.0-T MR imaging systems for clinical
From the University of Southern California, Los Angeles
applications is increasing worldwide. Importantly,
(F.G.S.), Keck School of Medicine and Institute for Magnetic
most previous investigations conducted to assess MR
Resonance, Safety, Education, and Research, Los Angeles, CA,
imaging safety of aneurysm clips used MR imaging
and the Divisions of Radiology (J.A.T., P.M.R., T.J.M.), Neurora-
diology Section, and Surgery (P.A.R.), Section of Neurosurgery,
systems with static magnetic fields of 1.5 T or less (7,
The Cleveland Clinic Foundation, Cleveland, OH.
11–25). In general, the increasing use of MR at 3.0 T
Address reprint requests to Frank G. Shellock, PhD, 7511 Mc-
requires additional studies to be performed to evalu-
Connell Avenue, Los Angeles, CA 90045.
ate metallic implants and devices at this field strength.
Thus, it is necessary to perform ex vivo testing at 3.0
T to characterize magnetic field–related safety for
by MR imaging system manufacturers), measurements of de-
aneurysm clips before allowing persons with these
flection angles may be substantially different. Therefore, in this
implants to enter this particular MR environment.
study, long- and short-bore MR imaging systems were used to
An important aspect of MR safety testing for me-
evaluate translational attraction for the aneurysm clips, as fol-
lows: long-bore MR system, actively shielded, head-only, MR
tallic implants involves the determination of magnetic
imaging system (length, 248 cm; bore inner diameter, 55 cm;
field interactions (ie, motion) and translational at-
3-T MR imaging system; General Electric Medical Systems,
traction (17, 18, 23–25). Translational attraction is
Milwaukee, WI); and short-bore MR system, actively shielded,
typically assessed by using the deflection angle test
head-only MR imaging system (length, 130 cm; bore inner
originally described by New et al (17), modified and
diameter, 60 cm; MAGNETOM, Allegra 3-T Headscanner;
used by others (18, 23–25), and recommended by the
Siemens Medical Systems, Erlangen, Germany).
American Society for Testing and Materials (27). Ac-
cording to this procedure, the deflection angle for an
Qualitative Evaluation of Magnetic Field Interaction
implant should be measured at the point of the high-
Each aneurysm clip was inspected at the entrance of the
est spatial gradient for the specific MR imaging sys-
imaging system bore to determine the qualitative presence of
tem used for testing (23–27). If the deflection angle
magnetic field interactions with the 3.0-T MR imaging systems.
from the vertical is less than 45 degrees, the implant
This was defined as any visual observance of directional move-
ment, rotation, or alignment to the magnetic field. The results
passes the translational attraction test insofar as the
were scored as either positive (observable motion, as de-
magnetic force acting on the implant is less than the
scribed) or negative (absolutely no motion). The entrance of
the MR imaging system bore was the position selected for this
Various types of magnets exist for commercially
assessment because it provided an easy and rapid site for this
available 3.0-T MR imaging systems. The magnet
evaluation and represented the closest position of the “MR
configurations include conventional long- and short-
environment” (ie, the immediate area relative to the MR im-
bore imaging units used for head-only and whole-
body clinical applications. Because of physical differ-
ences in the position and magnitude of the highest
Assessment of Translational Attraction
spatial gradient for different magnets, measurements
Translational attraction was assessed for each aneurysm clip
of deflection angles for implants by using long- versus
by using a standardized procedure known as the deflection angle
short-bore MR imaging systems may produce sub-
test according to guidelines provided by the American Society
for Testing and Materials (27). The aneurysm clip was attached
stantially different results. Therefore, the purpose of
to a special test fixture to measure the deflection angle in the
this investigation was to evaluate magnetic field in-
long- and short-bore MR imaging systems at the points of the
teractions and translational attraction for 32 different
highest spatial gradients (23–25, 27). The test fixture consists of
aneurysm clips in association with exposure to long-
a sturdy structure capable of holding the aneurysm clip in a
and short-bore 3.0-T MR imaging systems. Implica-
proper position without deflection of the test fixture. The test
tions of the results of this study for patients and other
fixture has a plastic protractor with 0-degree graduated mark-
ings. The protractor is rigidly mounted to the structure. The
persons with aneurysm clips regarding the 3.0-T en-
zero-degree indicator on the protractor was oriented vertically.
The test fixture has a plastic bubble level permanently affixed
to the top to ensure proper orientation in the MR imaging
The aneurysm clip was suspended from a thin, light-weight
string (weight, Ͻ1% of the weight of the implant) that was
attached at the 0-degree indicator position on the protractor.
Thirty-two different aneurysm clips from various manufac-
The length of the string was 20 cm, allowing the aneurysm clip
turers were evaluated in this investigation. Each aneurysm clip
to be suspended from the test fixture and to hang freely in
was representative of the manufactured finished version and
space. Sources of forced air movement within the respective
was not altered in any manner before testing. These aneurysm
3.0-T MR imaging system bores were shut off during the de-
clips were selected for this study because they represent various
types of clips made from nonferromagnetic or weakly ferro-
Measurements of deflection angles for the aneurysm clip
magnetic materials (eg, stainless steel alloy, Phynox, Elgiloy,
were obtained at the positions in the 3.0-T MR imaging systems
commercially pure titanium, titanium alloy) used for temporary
that produced the greatest magnetically induced deflections
or permanent treatment of aneurysms or arteriovenous mal-
(ie, the points of the highest spatial gradients) (23–25, 27). This
formations. The Table lists specific information regarding the
position was determined for each 3.0-T MR imaging system by
aneurysm clips (ie, the name, material, and manufacturer).
using gauss line plots provided by the manufacturer, measure-
ments, and visual inspection to identify the location where the
spatial magnetic field gradient was the highest. For the long-
bore 3.0-T MR imaging system, the highest spatial gradient
According to the American Society for Testing and Materi-
occurs at a position that is 96 cm from isocenter. The magnetic
als (27), translational attraction should be assessed for implants
spatial gradient at this position is 3.3 T/m. For the short-bore
at the point of the highest spatial gradient for the MR imaging
3.0-T MR imaging system, the highest spatial gradient occurs at
system used for testing. This is done to evaluate the magnet-
a position that is 78 cm from isocenter. The magnetic spatial
related force at an extreme or worst-case position for a metallic
gradient at this position is 5.25 T/m. The locations of the
object. As previously stated, there are various types of magnets
highest spatial gradients were marked by using tape to facilitate
used for 3.0-T MR imaging systems, including long- and short-
repeated measurements of deflection angles for the aneurysm
bore imaging units used for head-only and whole-body clinical
applications. Because there are physical differences in the po-
Thus, the test fixture was placed at the point of the highest
sition and magnitude of the highest spatial gradient for a given
spatial gradient for the long- and short-bore 3.0-T MR imaging
magnet (based on a review of technical specifications provided
systems. The aneurysm clip was held on the test fixture so that
Aneurysm clips: evaluation of magnetic field interactions and translational attraction using “long-bore” and “short-bore” 3.0-T MR imaging sys- tems
straight, 2-mm blade;stainless steel alloy;Zeppelin Chirugishe Instrumente,Pullach, Germany
straight, 6-mm blade;stainless steel alloy;Zeppelin Chirugishe Instrumente,Pullach, Germany
straight, 7-mm blade;stainless steel alloy;Zeppelin Chirugishe Instrumente,Pullach, Germany
straight, 5-mm blade;C.P. titanium;NMT Neurosciences,Duluth, Georgia
curved, 7-mm blade;C.P. titanium;NMT Neurosciences,Duluth, Georgia
straight, 9-mm blade;C.P. titanium;Elekta Instruments,Atlanta, Georgia
curved, 11-mm blade;C.P. titanium;NMT Neurosciences,Duluth, Georgia
straight, 11-mm blade;C.P. titanium;NMT Neurosciences,Duluth, Georgia
straight, 11-mm blade;C.P. titanium;NMT Neurosciences,Duluth, Georgia
Note.—LB indicates long-bore; Mag., magnetic; SB, short-bore; C.P., commercially pure. Continued
C.P. titanium;NMT Neurosciences,Duluth, Georgia
straight, 13-mm blade;C.P. titanium;Elekta Instruments,Atlanta, Georgia
bent, 7.5-mm blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
deflected type aneurysm clip forpermanent occlusion;
angled, 10-mm serrated blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
straight, 21-mm serrated blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
permanent occlusion;straight, 19-mm non-serrated blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
bent, 8-mm blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
curved, 6-mm blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
bent, 7-mm blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
Continued
straight, 7-mm blade;Elgiloy;Mizuho America, Inc.;Beverly, Massachusetts
45-degree angled, 19-mm serratedblade;titanium alloy;Mizuho America, Inc.;Beverly, Massachusetts
bayonet, 7-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 9-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 14-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
curved, 15.3-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 20-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
bayonet, 20-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
angled, 7-mm blade;Phynox;Aesculap, Inc.;Center Valley, Pennsylvania
Continued
straight, 7-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 9-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
titanium model FT758T;bayonet, 12-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 11-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
straight, 20-mm blade;titanium alloy;Aesculap, Inc.;Center Valley, Pennsylvania
the string was vertical and was then released. The deflection
to 18 degrees. Aneurysm clips made from commer-
angle for the aneurysm clip from the vertical direction to the
cially pure titanium and titanium alloy displayed no
nearest 0.5 degree was measured three times and averaged
translational attraction (n ϭ 15), whereas those made
from stainless steel alloy, Phynox, and Elgiloy dis-
played positive deflection angles (n ϭ 17).
Deflection angle measurements obtained for the aneurysm
Discussion
clips during exposure to the long-bore MR imaging system
were compared with those recorded during exposure to the
MR imaging procedures may be unsafe for patients
short-bore MR imaging system by using a Wilcoxon Signed
with certain implants made from ferromagnetic or
Rank Test (StatView; SAS Institute, Inc., Cary, NC).
conductive materials because of problems associated
with movement, heating, or induced electrical cur-
rents (5–8, 17, 30–36). Regarding aneurysm clips,
heating and induced currents are not of concern be-
The findings for magnetic field interactions and
cause of the physical size and shape of these relatively
translational attraction for the aneurysm clips ex-
small implants (17, 18, 30–35). Notably, MR imaging–
posed to the long- and short-bore MR imaging sys-
related heating and induced currents have been re-
tems are summarized in the Table. Seventeen of the
ported for only those implants or devices that have
32 aneurysm clips showed positive magnetic field in-
elongated configurations or that are electronically
teractions. Deflection angles for the aneurysm clips
activated (eg, neurostimulation systems, cardiac pace-
were significantly (P Ͻ .001) higher on the short-bore
makers, etc.) (6–8, 31–36). Therefore, from an MR
3.0-T MR imaging system compared with those re-
safety consideration, it is primarily important to de-
corded on the long-bore 3.0-T MR imaging system.
termine magnetic qualities for aneurysm clips before
On the long-bore MR imaging system, deflection an-
allowing patients or other persons with these objects
gles ranged from 0 to 16 degrees. On the short-bore
into the MR environment. Because most previous
MR imaging system, deflection angles ranged from 0
testing of aneurysm clips was conducted at 1.5 T, as
static magnetic fields of MR imaging systems increase
the aneurysm clips in this study because no standard
above this level, further investigations are necessary
currently exists for the quantification technique for
to characterize MR safety for these implants.
torque and no measurement value is available for use
From a magnetic field consideration, translational
to designate whether an aneurysm clip is unsafe.
attraction or torque may cause movement or dislodg-
Therefore, only aneurysm clips that exhibit no mag-
ment of a ferromagnetic implant, resulting in injury
netic field movements are considered to be safe from
(6–9, 15, 17, 18, 23–25, 29, 30). Translational attrac-
tion is proportional to the strength of the static mag-
Aneurysm clips come in a wide variety of shapes
netic field, the strength of the spatial gradient, the
and blade lengths and are made from different ma-
mass of the object, the shape of the object, and the
terials with varying magnetic susceptibilities. Each of
magnetic susceptibility of the object (17, 23–25, 29,
these factors can influence the MR safety aspects of
30). The effects of translational attraction on ferro-
these implants. In the present study, aneurysm clips
magnetic objects are predominantly responsible for
had shapes that included straight, bent, curved, and
possible hazards in the MR environment (ie, imme-
angled versions with blade lengths that ranged from 2
diate area around MR imaging system) (5, 8, 30). The
mm (Perneczky; Zeppelin Chirugishe Instrumente,
deflection angle test is commonly used to determine
Pullach, Germany) to 21 mm (Sugita, Large Aneu-
magnetic field–related translational attraction for im-
rysm Clip for Permanent Occlusion; Mizuho Amer-
plants, materials, and devices (17, 18, 23–25, 29).
ica, Inc., Beverly, MA). Materials used to make these
The American Society for Testing and Materials
aneurysm clips included stainless steel alloy, Phynox,
guidelines for deflection angle testing of implants in
Elgiloy, commercially pure titanium, and titanium
the MR environment, indicate that, “. . . if the im-
plant deflects less than 45 degrees, then the magnet-
Previous reports investigating magnetic qualities of
ically induced deflection force is less than the force on
aneurysm clips indicated that every aneurysm clip
the implant due to gravity (its weight)” (27). For this
made from stainless steel alloy, Phynox, Elgiloy, com-
condition, it is assumed that any risk imposed by the
mercially pure titanium, and titanium alloy was safe at
application of the magnetically induced force is no
1.5 T (6–8, 11–14, 15–26). In consideration of the
greater than any risk imposed by normal daily activity
current knowledge pertaining to aneurysm clips at 1.5
in the earth’s gravitational field (27). Accordingly,
T, the following guidelines have been recommended
findings from the deflection angle test permit im-
for careful consideration before performing MR im-
plants and devices made from nonferromagnetic or
aging in a patient with an aneurysm clip and before
weakly ferromagnetic materials that display deflec-
allowing any person with an aneurysm clip into the
tion angles between 0 and 44 degrees to be present in
patients or other persons in the MR environment
Torque, which tends to align a ferromagnetic object
parallel to the magnetic field, is dependent on the
Guidelines Regarding Aneurysm Clips and the
strength of the magnetic field, the dimensions of the
object, and the initial angulation of the object relative
1. Specific information (ie, manufacturer, type or
to the static magnetic field (17, 23–25, 30). Torque
model, material, lot and serial numbers) regarding
effects on ferromagnetic objects are mainly responsi-
the aneurysm clip must be known, especially with
ble for possible hazards during an MR imaging pro-
respect to the material used to make the aneurysm
cedure, when the patient is positioned at the center of
clip, so that only patients or other persons with non-
the MR imaging system (ie, the position where torque
ferromagnetic or weakly ferromagnetic clips are al-
lowed into the MR environment. The manufacturer
A variety of techniques have been used to qualita-
provides this information in the labeling of every
tively or quantitatively determine magnetic field-re-
aneurysm clip. The implanting surgeon is responsible
lated torque for implants and devices (17, 23, 25, 29,
for properly communicating this information in the
30). To date, a test procedure and acceptable mea-
surement value for torque imposed on implants has
2. An aneurysm clip that is in its original package
not been defined by the American Society for Testing
and made from Phynox, Elgiloy, MP35N, titanium
and Materials. However, according to the American
alloy, commercially pure titanium, or other material
Society for Testing and Materials (27), a torque value
known to be nonferromagnetic or weakly ferromag-
for an implant “that is less than that produced by
netic does not need to be evaluated for ferromag-
normal daily activities (which might include rapidly
netism. Aneurysm clips made from nonferromagnetic
accelerating vehicles or amusement park rides) is
or weakly ferromagnetic materials in original pack-
assumed to be safe.” Notably, the amount of torque
ages do not require testing of ferromagnetism be-
necessary to displace an aneurysm clip is unknown,
cause the manufacturers ensure the pertinent MR
particularly because counter forces (eg, related to the
safety aspects of these clips and, therefore, should be
closing force of the clip, granulation of tissue, and
held responsible for the accuracy of the labeling.
other factors) may be present that require additional
3. If the aneurysm clip is not in its original package
characterization, possibly by using in vivo techniques.
and properly labeled, it should undergo testing for
Therefore, torque was not specifically determined for
4. The radiologist and implanting surgeon should
rysm clips made from commercially pure titanium or
be responsible for evaluating the available informa-
titanium alloy seem to be entirely safe because of the
tion pertaining to the aneurysm clip, verifying its
total lack of magnet-related movements. The remain-
accuracy, obtaining written documentation and decid-
ing aneurysm clips made from stainless steel alloy,
ing to perform the MR procedure after considering
Phynox, and Elgiloy require characterization of mag-
the risks versus the benefits of the examination.
netic field-induced torque to determine whether they
Of note is that Brothers et al (37) evaluated pa-
are safe for patients during MR imaging procedures.
tients after surgery for vertebrobasilar aneurysms
Notably, these results are specific to the 3.0-T MR
with nonferromagnetic Sugita aneurysm clips at 1.5 T
imaging systems used for this evaluation or with com-
and reported that no ill effects occurred. In addition,
Pride et al (26) conducted a study of patients with
nonferromagnetic aneurysm clips who underwent
MR imaging. No objective adverse outcome occurred
Long-Bore versus Short-Bore Deflection
in these patients, further confirming that MR imaging
can be performed safely in patients with nonferro-
An interesting finding of this study is that there
were significantly (P Ͻ .001) higher deflection angles
However, as previously discussed, few studies have
measured for the aneurysm clips during exposure to
been performed to evaluate magnetic field interac-
the short-bore versus the long-bore 3.0-T MR system.
tions of implants in association with MR imaging
To our knowledge, this is the first description of such
systems operating above 1.5 T (28, 29). A study con-
an important phenomenon that is obviously due to
ducted at 8.0 T by Kangarlu and Shellock (29) re-
the higher spatial gradient associated with the short-
ported that all aneurysm clips, even those made from
bore imaging unit. Although this did not impact the
titanium or titanium alloy, displayed positive transla-
MR imaging safety aspects of the aneurysm clips in
tional attractions (deflection angles ranged from 5 to
this study, it is conceivable that other metallic im-
53 degrees). Importantly, several aneurysm clips re-
plants may be found to be safe on the long-bore MR
ported to be safe at 1.5 T (6–8, 17, 18, 23) were found
imaging system and unsafe on a short-bore MR im-
to be potentially unsafe at 8.0 T because they showed
aging system. Therefore, further study of this issue is
excessive deflection angles and relatively high quali-
tative torque values (29). In view of the findings at 8.0
T and because of the proliferation of 3.0-T MR im-
aging systems, it was considered important to deter-
Acknowledgments
mine magnetic field–related safety for comparable
Special thanks to Dr. Mark Cohen, Dr. Susan Bookheimer,
and Dr. John Mazziotta at the University of California, Los
Findings from the present study indicated that only
Angeles Brain Mapping Division, Los Angeles, CA, for per-
the aneurysm clips made from commercially pure
mitting use of the 3.0-T MR imaging system and, thus, facili-
titanium or titanium alloy are definitely safe because
tating the performance of this research project.
they exhibit no magnet-related movements in associ-
ation with exposure to 3.0-T MR imaging systems. References
Aneurysm clips made from stainless steel alloy,
Phynox, and Elgiloy, while displaying acceptable de-
1. Yasargil MG. Microsurgery Applied to Neurosurgery. Stuttgart:
flection angles (Ͻ45 degrees) and thus considered
2. Dujovny M, Kossovsky N, Kossovsky R, et al. Intracranial clips: an
safe for patients and other persons in the long- and
examination of the devices used for aneurysm surgery. Neurosurgery
short-bore MR environments (again, the immediate
areas associated with the MR imaging systems up to
3. Dujovny M, Kossovsky, Kossowsky R, et al. Vascular clips: an historic and biomedical perspective. In: Fein JN, Flamm E, eds.
and including the entrances of the magnet bores),
Cerebral Vascular Surgery. New York: Springer-Verlag; 1985:997–
require further characterization of torque effects to
determine safety for patients who have these clips
4. Mayfield FH, Kees G Jr. A brief history of the development of the Mayfield clip: technical note. J Neurosurg 1971;35:97–100
before allowing them to undergo MR imaging proce-
5. Sawyer-Glover A, Shellock FG. Pre-MRI procedure screening: rec- ommendations and safety considerations for biomedical implants
Thus, from a practical consideration, the results of
and devices. J Magn Reson Imaging 2000;12:92–106
this investigation have implications for two different
6. Shellock FG. Magnetic Resonance Procedures: Health Effects andSafety. Boca Raton: CRC Press; 2001
situations. First, regarding the long- and short-bore
7. Shellock FG. Guide to MR Procedures and Metallic Objects: Update
3.0-T MR environments, all aneurysm clips that were
2001. 7th ed. Philadelphia: Lippincott Williams & Wilkins; 2001
assessed seem to be safe because of the relatively
8. Shellock FG. Reference Manual for Magnetic Resonance Safety: 2003Edition. Salt Lake City: Amirsys, Inc.; 2003
minor magnetic field-related translational attractions
9. Klucznik RP, Carrier DA, Pyka R, Haid RW. Placement of a
that were measured (deflection angles Ͻ45 degrees). ferromagnetic intracerebral aneurysm clip in a magnetic field with
Therefore, patients and other persons (eg, MR tech-
a fatal outcome. Radiology 1993;187:855– 856
nologist, family member, etc.) with these specific an-
Need for caution during MR imaging of patients with aneurysm clips. Radiology 1993;188:287–288
eurysm clips would be permitted into the respective
11. Becker R, Norfray J, Teitelbaum G, Bradley WG Jr, Jacobs JB,
3.0-T MR environments. Second, for patients under-
Wacaser L, Rieman RL. MR imaging in patients with intracranial
going MR imaging procedures with the use of long- or
aneurysm clips. AJNR Am J Neuroradiol 1988;9:885– 889
12. Dujovny M, Kossovsky N, Kossowsky R, et al. Aneurysm clip
short-bore 3.0-T MR imaging systems, only the aneu-
motion during magnetic resonance imaging: in vivo experimental study with metallurgical factor analysis. Neurosurgery 1985;17:543–
27. American Society for Testing and Materials (ASTM) Designation:
F 2052. Standard test method for measurement of magnetically
13. Dujovny M, Alp MS, Slavin KV, et al. Magnetic characteristics of
induced displacement force on passive implants in the magnetic
Yasargil aneurysm clips. Surg Neurol 1997;47:547–550
resonance environment. In: Annual Book of ASTM Standards,
14. Dujovny M, Gundamjaj NR, Misra M, Alp MS. Aneurysm clips.
Section 13, Medical Devices and Services, Volume 13.01 Medical
Crit Rev Neurosurg 1997;7:169–175
Devices, Emergency Medical Services. West Conshohocken, PA,
15. Kanal E, Shellock FG, Lewin JS. Aneurysm clip testing for ferro- magnetic properties: clip variability issues. Radiology 1996;200:
28. Hennemeyer CT, Wicklow K, Feinberg DA, Derdeyn CP. In vitro evaluation of platinum Guglielmi detachable coils at 3 T with a
16. Romner B, Olsson M, Ljunggren B, et al. Magnetic resonance porcine model: safety issues and artifacts. Radiology 2001;219:732– imaging and aneurysm clips. J Neurosurg 1989;70:426 – 431
17. New PF, Rosen BR, Brady TJ, et al. Potential hazards and artifacts
29. Kangarlu A, Shellock FG. Aneurysm clips: evaluation of magnetic of ferromagnetic and non-ferromagnetic surgical and dental mate- field interactions with an 8.0 T MR system. J Magn Reson Imaging rials and aneurysm clips in nuclear magnetic resonance imaging.
30. Schenck JF. Health Effects and Safety of Static Magnetic Fields. In:
18. Shellock FG, Crues J. High-field strength MR imaging and metal- Magnetic Resonance Procedures: Health Effects and Safety. Boca
lic biomedical implants: an ex vivo evaluation of deflection forces. AJR Am J Roentgenol 1988;151:389–392
31. Davis PL, Crooks L, Arakawa M, et al.
19. Kato Y, Sano H, Katada K, et al. Potential hazards in NMR Effects of new titanium cerebral imaging: heating effects of changing magnetic fields and RF fields aneurysm clips on MRI and CT images. Minim Invasive Neurosurg on small metallic implants. AJR Am J Roentgenol 1981;137:857– 860
20. Lawton MT, Heiserman JE, Prendergast VC, Zabramski JM, Spet-
32. Shellock FG. Radiofrequency-induced heating during MR proce-
zler RF. Titanium aneurysm clips: part III. clinical application in dures: a review. J Magn Reson Imag 2000;12:30 –36 16 patients with subarachnoid hemorrhage. Neurosurgery 1996;38:
33. Smith CD, Kildishev AV, Nyenhuis JA, Foster KS, Bourland JD. Interactions of MRI magnetic fields with elongated medical im-
21. Piepgras A, Guckel F, Weik T, Schmiedek P. Titanium aneurysm plants. J Appl Phys 2000;87:6188 – 6190 clips and their advantages in diagnostic imaging [in German].
34. Nyehnuis JA, Kildishev AV, Foster KS, Graber G, Athey W. Heat- ing near implanted medical devices by the MRI RF-magnetic field.
22. Wichmann W, Von Ammon K, Fink U, Weik T, Yasargil GM. IEEE Trans Magn 1999;35:4133–4135
Aneurysm clips made of titanium: characteristics and artifacts in
35. Smith CD, Nyenhuis JA, Kildishev AV. Health effects of induced MR. AJNR Am J Neuroradiol 1997;18:939 –944 electrical currents: implications for implants. In: Shellock FG, ed.
23. Shellock FG, Kanal E. Yasargil aneurysm clips: evaluation of Magnetic Resonance: Health Effects and Safety. Boca Raton: CRC
interactions with a 1.5-T MR system. Radiology 1998;207:587–591
24. Kanal E, Shellock FG. Aneurysm clips: effects of long-term and multiple
36. Rezai AR, Finelli D, Nyenhuis JA, et al. Neurostimulator for deep exposures to a 1.5-T MR system. Radiology 1999;210:563–565 brain stimulation: in vitro evaluation of magnetic resonance imag-
25. Shellock FG, Shellock VJ. Spetzler titanium aneurysm clips: com- ing-related heating at 1.5 tesla. J Magn Reson Imaging 2002;15:141– patibility at MR imaging. Radiology 1998;206:838 – 841
26. Pride GL, Kowal J, Mendelsohn DB, Chason DP, Fleckenstein JL.
37. Brothers MF, Fox AJ, Lee DH, Pelz DM, Deveikis JP. MR imaging Safety of MR scanning in patients with nonferromagnetic aneu- after surgery for vertebrobasilar aneurysm. AJNR Am J Neurora- rysm clips. J Magn Reson Imaging 2000;12:198 –200
1329 East 13th. Ave., Vancouver, B.C. V5N 2B5 Suite 600 – 890 West Pender St., Vancouver, B.C. V6C 1J9 Hours 8am to 8pm 7 Days Per Week for Drop Off Only Dealer Code: __________________________ Dealer Name: ______________________________ Name: ________________________________ Company Name____ _________________ _____ Email: _______________________________________ Tel : ______________
A Comparison of 10 and 14 Days of Lansoprazole Triple Therapy for Eradication of Helicobacter pylori M. Brian Fennerty, MD; T. O. G. Kovacs, MD; R. Krause, MD; M. Haber, MD;A. Weissfeld, MD; N. Siepman, MD; P. Rose, MD Background: Data from large, multicenter, US studies in 84% (103/123) of those receiving 10-day triple therapydetermining the efficacy of triple therapy for the eradi-by